42

ZOLL AED Plus Guida per l’amministratore

Tabella 12: Risultati prestazioni cliniche (pazienti pediatrici)

*155 delle 415 registrazioni di ritmi anomali erano SVT (39 pazienti).

1. Le prestazioni in caso di aritmia sono tratte dall’articolo: RE Kerber, LB Becker, JD Bourland, RO Cummins,

AP Hallstrom, MB Michos, G Nichol, JP Ornato, WH Thies, RD White, BD Zuckerman, “Automated External
Defibrillators for Public Access Defibrillation: Recommendations for Specifying and Reporting Arrhythmia
Analysis Algorithm Performance, Incorporation New Waveforms, and Enhancing Safety”, Circulation 1997,
Vol 95, No 6, 1677-1681

Riferimenti bibliografici:

Young KD, Lewis RJ: “What is confidence? Part 2: Detailed definition and determination of confidence
intervals”. Annals of Emergency Medicine, September 1997; 30; 311-218

“CRC Standard Mathematical Tables 28

th

Edition”, William H. Beyer, Ph.D., CRC Press, Inc,

Boca Raton, FL., 1981, Percentage Points, F-Distribution Table, pg 573.

Ritmi

Dimensioni

campione

(registrazioni di

9 secondi)

Obbiettivi

prestazioni

Prestazioni

osservate

Limite minimo di

confidenza da un

solo lato, 90%

Defibrillabile (23 pazienti)

Fibrillazione ventricolare grossolana

(1)

Tachicardia ventricolare rapida

(2)

42
51

Sensibilità >90%
Sensibilità >75%

100% (42/42)
92.2% (47/51)

93.1%
82.9%

Non defibrillabile (121 pazienti)

Ritmo sinusale normale,

NSR (3)

AF, SB, SVT*, blocco cardiaco,
idioventricolare, PVCs (4)
Asistolia (5)

229
415

14

Specificità >99%
Specificità >95%

Specificità >95%

100% (229/229)
100% (415/415)

100% (14/14)

98.7%
99.3%

80.7%

Intermedio (15 pazienti)

FV fine (6)

Altre tachicardie ventricolari

(7)

0
22

Solo rapporto
Solo rapporto

n.p.
77.3% (17/22)

---
58.0%

Advertising