ZOLL AED Pro Rev F Manuale d'uso
Pagina 102

A
PPENDICE
A S
PECIFICHE
A–16
www.zoll.com
9650-0350-11 Rev. F
Le prestazioni in caso di aritmia sono tratte dall’articolo: Kerber RE, Becker LB, Bourland JD,
Cummins RO, Hallstrom AP, Michos MB, Nichol G, Ornato JP, Thies WH, White RD, Zuckerman
BD. “Automated External Defibrillators for Public Access Defibrillation: Recommendations for
Specifying and Reporting Arrhythmia Analysis Algorithm Performance, Incorporating New
Waveforms, and Enhancing Safety,” Circ J Am Heart Assoc. 1997;95:1677-1682.
Riferimenti bibliografici
Young KD, Lewis RJ. “What is confidence? Part 2: Detailed definition and determination
of confidence intervals”. Ann Emerg Med. September 1997;30:311-318.
Beyer WH. Percentage Points, F-Distribution Table. CRC Standard Mathematical Tables.
28
th
ed. Boca Raton, Fla: CRC Press; 1981:573.
Tabella A-5.
Risultati prestazioni cliniche (pazienti pediatrici)
Ritmi
Dimensioni
campione
(registrazioni di
9 secondi)
Obbiettivi
prestazioni
Prestazioni
osservate
Limite minimo di
confidenza da un
solo lato, 90%
Defibrillabile
(49 pazienti)
Sensibilità
Fibrillazione
ventricolare
grossolana
42
>90%
100% (42/42)
93,1%
Tachicardia
ventricolare rapida
82
>75%
93,9% (77/82)
87,6%
Non defibrillabile
(155 pazienti)
Specificità
Ritmo sinusale
normale (NSR)
208
>99%
100% (208/208)
98,6%
Fibrillazione atriale,
bradicardia sinusale,
tachicardia
sopraventricolare
a
,
blocco cardiaco,
idioventricolare,
extrasistoli
ventricolari
a. 161 delle 348 registrazioni di ritmi anomali erano SVT (72 pazienti). Le frequenze cardiache
SVT erano comprese tra 152 e 302 battiti al minuto.
348
>95%
99,4% (346/348)
98,2%
Asistolia
29
>95%
100% (29/29)
90,2%
Intermedio
(16 pazienti)
FV fine
0
Solo
rapporto
—
—
Altre TV
40
Solo
rapporto
90% (36/40)
78,6%