8-15

USING THE SYSTEM

UK

8.5.1.6 POWER

SUPPLY

FAULTS

Power supply and display memory diagnostics

From the FAULTS panel (point 8.5.1), press numerical key 4 to
access the panel for viewing the diagnostics of the primary and
secondary power supplies and of the memory for managing the
display.

The diagnostic state of each item monitored is reported, as
shown in the following table.

Press ESC to return to the FAULTS panel.

Label

Part being diagnosed

Display

Cause

Action required

All

All

OK

No failures

-

N.T.

NOT TEST. Diagnostics disabled
during configuration (See Set>
BACKGROUND TEST menu)

-

Res.

Failure cleared due to previous
failure signal

Access the Report> FAULTS panel from the
OPERATOR menu and carry out a MANUAL
RESET of the FAILURE signal.

Main P.S.

Primary 230 VAC
power supply

Fault

Primary power supply
not detected

Check the 230 VAC power supply, the connecting
cable and the mains fuse.
If the failure persists, call Servicing.

24Vdc P.S.

Secondary 24 VDC
power supply

Secondary power supply
not detected

Check the external 24 VDC power supply system.
If the failure persists, call Servicing.

Display

Memory managing the
graphic display

Memory checksum error

Contact Servicing.