Bird Model 6091 Calorimeter

42

Status Words

(Ux)

Function

Read one status word and return information in character string.

Set x to:
"0" for Machine Status

"1" for Error Status

"2" for Revision History

Machine Status Word (U0) - Contents include current operating parameters. To
ensure correct status is issued, this status word should be requested as soon as
possible after command is sent. Operator may choose to record this data for re-
configuration of device to previous parameters.

Machine Status Word:

1 2 3 4 5 6 7

8 9

10 11 12 13 14 15

16 17

18 19 20

21

-

6 0 9 1 -

W A P

Y

Y

T

T

1

M

0

0

K

Y

CR LF

Terminators per bits 12 &
13
EIO (KY, KN)
SRQ Mask (M00 to M63)
Trigger (T1 through T5)
Terminator (YT, YO, YN)
Prefix (PY, PN)
Measurement Command