Chapter 5

Maintenance

With basic care, the Calorimeter should provide many years of trouble free operation.

This chapter contains operator maintenance, performance testing, preventive main-

tenance, and cleaning instructions.

Performance Testing

Check for proper equipment operation using the performance test procedures in this

section. Acceptable performance is determined by comparing the test results with the

equipment specifications listed in Chapter 1 - Introduction. If the unit does not pass

the performance test, notify Bird Electronic Corporation.

Equipment Required

Refer to table 3 for a listing of the test equipment required for performance testing.

The equipment parameters listed are the minimum specifications applicable to the

test equipment models other than those listed. Valid Calorimeter tests and measure-

ments are critical and depend largely on the accuracy of the test equipment used.

Test Data Sheet

A test data sheet is provided on page 21. Photocopy the data sheet. Record perfor-

mance test data in the spaces provided. Maintain completed test data sheets for re-

cords of equipment performance and testing.

Performance Test

Setup

Set up the Calorimeter and test equipment by following the instructions below. Refer

to figure 12.

1. Connect the cable assembly from the Calorimeter to the ac power me-

ter.

2. Connect the power meter to the ac power source.

Calorimeter

Performance Test

Perform the calorimeter performance testing as follows:

CAUTION

Do not apply RF power to the load without proper coolant flow. Failure to keep coolant at

proper flow rate can result in serious damage to the equipment.

1. Turn power on to Calorimeter and test equipment. Allow a 30 minute

warm-up period for all electronic equipment. While the equipment is

19

Equipment Type

Suggested Model

Critical Equipment Specifications

AC Power Source Elgar 501SL with PIP

9012

Output - 10 to 200 watts @ 50 ohms
Stability - greater than 0.1%
THD - less than 0.25%
Frequency - adjustable to 60, ±1 Hz

AC Power Meter

YEW 2533

Range - 10 to 200 watts
Accuracy - 0.1%

Automatic Net-
work Analyzer

HP 8753 with 85046A

±0.2 dB accuracy for return loss
measurements

Cable Assembly

Bird P/N 6091-099

Test connections as shown in figure
12

Table 3

Test

Equipment

Specifications