WARNING

Do not place compressed air near or directly against skin. Do not use compressed air in excess of 30 psi (207 kPa)

when cleaning or drying parts. Improper use of compressed air can cause serious injury or death.

WARNING

Coolant containing ethylene glycol is a potential health hazard. Avoid ingestion, inhaling of vapors and eye and skin

contact.

WARNING

Do not attempt to handle or move this unit alone. More than one person is needed to move this unit to avoid

possible injury.

Caution Statements

The following equipment cautions appear in the text whenever the equipment is in danger of damage, and

are repeated here for emphasis.

CAUTION

The Calorimeter can overheat if operated without sufficient coolant. Failure to keep coolant reservoir adequately filled

can result in serious damage to the equipment.

CAUTION

Use only premixed coolant Bird part number 6091-120. Use of other solutions will damage the instrument and void

all warranties.

CAUTION

Do not tighten the jack screws with a screwdriver. The screwdriver slots in the screw are provided for removal

purposes only.

CAUTION

During remote operation, periodically monitor the bus service request line. Failure to detect the service request can

result in major equipment damage.

CAUTION

The rack enclosure must provide at least 100 ft

3

/min of unrestricted air flow. Avoid mounting the Calorimeter above

instruments that give off heat. If necessary, isolate the Calorimeter from heat-generating devices with foam or

fiberglass.

CAUTION

The total length of all cables used must be less than the number of devices connected times 2 meters (6.6 feet).

Never exceed 20 meters of total length.

CAUTION

Only finger-tighten the GPIB connector screws. The screwdriver slots are provided for loosening and removal of

screws only. Connector screw threads may be stripped if tightened with a screwdriver.

CAUTION

Never turn off the Calorimeter’s main power switch while RF source power is applied to Calorimeter input. Failure

to turn off RF power source can result in major equipment damage.

CAUTION

Do not apply RF power to the load without proper coolant flow. Failure to keep coolant at proper flow rate can

result in serious damage to the equipment.