Bird Technologies 6091P Manuale d'uso

Pagina 52

Advertising
background image

Bird Model 6091 Calorimeter

40

Prefixes

(Px)

Function

Turns the prefix mode on or off.

Set x to:

"Y" for enabling prefix messages.

"N" for suppressing prefix messages.

Remarks

Prefixes indicate status of current measurement.

"N" (Normal) prefix when reading is within ±3% of final value.

"Y" (Transient prefix when reading is not stable.

On power up or after DCL/SDC, default is prefix mode on "PY".

Statements

National GPIB-PCII

HP 200/300

LET CMD$ = "PY"

OUTPUT 724; "PY"

CALL IBWRT (UD%,CMD$)

Triggers

(Tx)

Function

Selects the type of stimulus on which the instrument will trigger to initiate a measure-
ment.

Set x to:

"0" for continuous on Talk.

"1" for one shot on Talk.

"2" for continuous on GET.

"3" for one shot on GET.

"4" for continuous on measurement command. (WA, FL, IN, OU or DT)

"5" for one shot on continuous measurement command. (WA, FL, IN, OU or DT)

Remarks

Failure to trigger device before requesting a reading will lock up the bus.

T0 or T1 trigger. The measurement is triggered when the Calorimeter is addressed to
talk.

T1 trigger. Halts bus until reading is available.

T4 and T5 trigger. Reading is triggered only after receiving a measurement command.

On power up or following DCL/SDC, the default trigger mode is one shot measure-
ment on Talk (T1).

Statements

National GPIB-PCII

HP 200/300

LET CMD$="TO"

OUTPUT 724; "T0"

CALL IBWRT (UD%,CMD$)

Advertising