Re mote op er ating pro ce dures. 31, Gpib over view. 31, Gpib ca pa bil ities. 31 – Bird Technologies 6091P Manuale d'uso

Pagina 41: Remote operating procedures, Gpib overview, Gpib capabilities

Advertising
background image

Only those functions within the scope of normal maintenance are listed. This manual

cannot list all malfunctions that may occur, or corrective actions. If a malfunction is

not listed or is not corrected by listed corrective actions, notify a qualified service cen-

ter.

Preparation for Shipment or Storage

The following section provides the special instruction needed to prepare the instru-

ment for shipment or storage.

Storage

Requirements

The Calorimeter should be stored in a cool, dry area. Ambient temperature of the

storage area must be within °10 to 50°C (50° to 122°F). If storage period is expected

to exceed 30 days, drain coolant per the following paragraph. Package the unit per

the Packaging section to keep it free of dust and dirt, and to protect it against rough

handling.

Draining the

Coolant

Prior to storing or shipping the Calorimeter, drain the coolant according to the fol-

lowing procedure:

WARNING

Coolant containing ethylene glycol is a potential health hazard. Avoid ingestion, inhaling of

vapors and eye and skin contact.

1. Remove four screws securing the top cover to the instrument case, if

used. Refer to figure 18.

2. Place the Calorimeter unit on a flat surface with the rear panel ex-

tending out over the edge.

3. Place a suitable container below the drain tube. With the container in

place, remove the reservoir filler cap and turn the hex on the drain

valve counterclockwise to open the drain. Allow all of the coolant to

drain from the Calorimeter.

29

Maintenance

Figure 18

Coolant Draining