Specifications

Power Range

10 to 200 watts (average power, independent of wave
shape)

Frequency Range

DC to 2500 MHz

Accuracy of Reading

1

:

@ 10 to 25 watts
@ 25 to 200 watts

±3%
±1.25%

Input Connector

Female, Type N

Input Impedance

50 ohms

Input VSWR:

DC to 1000 MHz
1000 to 2500 MHz

1.10 maximum (26.4 dB minimum return loss)
1.25 maximum (19.1 dB minimum return loss)

Response Time

1

(10 to 200 watt step response)

1 minute maximum to reach 97% of final reading

GPIB Function Supported
(IEEE 488, 1975)

SH1, AH1, T6, L4, SR1, RL1, PP0, DC1, CO, E1, TEO,
and LE0

Maximum Bus Reading Rate

3 readings per second

Power Requirements (Single phase)

104 to 132 Vac or 195 to 264 Vac @ 47 to 63 Hz

Power Consumption

45 to 70 watts

Coolant Type

Ethylene Glycol mixture (Bird part number 6091-120)

Coolant Capacity

525 ml (10 fluid oz)

Operating Position

Horizontal

Storage Temperature

10°C to 50°C (50°F to 122°F)

Operating Temperature

1

15°C to 35°C (59°F to 95°F)

Weight:

Model 6091
Model 6091P

18.1 kg (40 lb)
11.3 kg (25 lg)

Dimensions:

Model 6091

Model 6091P

15-3/4”L x 20-15/32”W x 8-17/64”H
(400 mm x 520 mm x 210 mm)
14-7/8”L x 19”W x 6-31/32”H
(378mm x 483 mm x 177 mm)

1

To achieve the accuracy and response time specified, the Calorimeter must be operated in a stable ±5°C

(±9 F°) ambient temperature environment. Accuracy specification does not include losses due to mismatch

.