Chapter 3

Installation

This chapter contains information on site and shelter requirements, unpacking and

inspection, and preparing the Calorimeter for use.

Site and Shelter

Requirements

The Model 6091/6091P Calorimeter is designed for use in a laboratory environment.

The unit is not intended for outdoor use, or use in areas of condensing humidity. It

must be located near an ac power source.

WARNING

Do not attempt to handle or move this unit alone. More than one person is needed to move

this unit to avoid possible injury.

Unpacking and

Inspecting

1. Carefully inspect the Calorimeter shipping container for signs of

damage. If damage is noticed, do not unpack the unit. Immediately

notify the shipping carrier and Bird Electronics of the damage.

2. If the container is not damaged, unpack the unit. Save the shipping

materials for repackaging.

3. Check contents against the equipment list in Chapter 1. Inspect all

components for visual signs of damage. Immediately notify the ship-

ping carrier and Bird Electronic Corporation of equipment damage or

missing parts.

Setting the Voltage

Selector Switch

The Calorimeter can be operated at 104 to 132 volts RMS or 195 to 264 volts RMS.

The sensor unit is preset at the factory for 115 Vac operation. Verify voltage selection

and fuse rating; follow the steps below if a change is required.

1. Locate the power entry module, refer to figure 5, on the back of the Calorime-

ter unit.

7

100V-120V

200V-240V

100V-120V

WITH

250V

USE

ONL

Y

FUSES

Module

ON/OFF

Switch

Fuse

Fuse

Voltage

Selector Card

Fuse Drawer

Figure 5

Power Entry

Module