Pagina 20: Connecting gpib cables

Advertising
background image

2. Use a small flathead screwdriver to gently pry the fuse drawer out of the

power entry module.

3. Slide the voltage selector card out of the fuse drawer. Reinsert the card so that

the correct line voltage appears in the window.

4. Check the current rating of the fuses with requirements in table 2. Install the

correctly rated fuses in the fuse drawer.

5. Re-install the fuse drawer in the power entry module.

AC Power

Connection

An ac power cord is supplied with the Calorimeter unit. The female end of the power

cord plugs into the power entry module, refer to figure 5.

The male end of the power cord is equipped with a U.S. standard NEMA 5-15,

three-pronged connector for mating with a grounded 120-volt wall receptacle. You

must purchase the appropriate cord for connection to any other receptacle.

Make ac power connections as follows:

1. Locate the power ON/OFF switch on the rear panel of the unit. Place the

switch in the OFF position.

2. Check the input voltage rating indicated on the power entry module. Verify

that the voltage indicated matches the supply voltage for your installation. If

not, change the voltage selector switch per the Setting the Voltage Selector

Switch section.

3. Plug the female end of the power cord into the power entry module.

4. Plug the male end of the power cord into a mating three-pronged receptacle.

Filling the Coolant

Reservoir

The Calorimeter requires 525 ml (10 fl oz) of coolant (Bird P/N 6091-120). The coolant

reservoir is emptied prior to shipment or extended storage. Refer to figure 6 and fill

the coolant reservoir as follows:

WARNING

Coolant containing ethylene glycol is a potential health hazard. Avoid ingestion, inhaling of

vapors and eye and skin contact.

CAUTION

Use only premixed coolant Bird part number 6091-120. Use of other solutions will damage

the instrument and void all warranties.

CAUTION

The Calorimeter can overheat if operated without sufficient coolant. Failure to keep coolant

reservoir adequately filled can result in serious damage to the equipment.

CAUTION

Do not apply RF power to the load without proper coolant flow. Failure to keep coolant at

proper flow rate can result in serious damage to the equipment.

Bird Model 6091 Calorimeter

8

Voltage Setting

Voltage Range

(volts RMS)

*Fuse Rating

Bird Part No.

100-120V

104 to 132

2.0A, 250V Fast Acting

5-1976-19

220-240V

195 to 264

1.0A, 250V Fast Acting

5-1976-16

* Replace with only IEC 5 x 20 mm type fuses.

Table 2

Fuse Ratings

and Part

Numbers