Main te nance. 19, Per for mance testing. 19, Equip ment re quired. 19 – Bird Technologies 6091P Manuale d'uso

Pagina 29: Test data sheet. 19, Per for mance test setup. 19, Cal o rim e ter per for mance test. 19, Maintenance, Performance testing, Equipment required, Test data sheet

Advertising
background image

Typical response times for the Calorimeter are shown in figure 11. The first curve

shows the response time with power stepped up from 0 to 200 watts. The second

curve indicates the response when the input is stepped down from 200 to 0 watts.

The step-down response time is considerably longer than the step-up response time.

This is due to residual heat stored in the system.

Measuring Coolant

Flow Rate

The flow rate is displayed in liters per minute. The flow rate should be between 0.284

to 0.473 liters per minute. An alarm will sound if the flow rate drifts outside of this

range.

1. Press and hold the FLOW pushbutton to check the coolant flow rate.

2. Release the FLOW pushbutton to return the display to the power dis-

play mode.

Measuring

Temperature

Differential (

DT)

The temperature differential is displayed in degrees Celsius. The

T is normally be-

tween 0.380° to 7.610°C and increases linearly with the RF power input level. An

alarm will sound if the

T exceeds 8.500°C. If the alarm sounds, shut down according

to the emergency shutdown procedures in this chapter.

1. Press the

T pushbutton to check the rise in temperature as the cool-

ant passes through the load resistor.

2. Release the

T pushbutton to return the display to the power display

mode.

GPIB Operation

The Calorimeter is capable of remote operation via the General Purpose Interface

Bus (GPIB). Detailed information relating to bus capabilities together with program-

ming examples are detailed in Appendix A - Remote Operating Procedures.

Normal Shutdown Procedure

When power measurements are completed, shutdown the Calorimeter as follows:

CAUTION

Never turn off the Calorimeter’s main power switch while RF source power is applied to

Calorimeter input. Failure to turn off RF power source can result in major equipment

damage.

1. Turn off RF power source device.

17

Operating Instructions

Figure 11

Typical

Response Times

Advertising
Questo manuale è associato alle seguenti prodotti: