The contact assignment of the 24-pin GPIB cable connector is shown in figure 7.

Setting the GPIB Bus Address

Each instrument connected to the GPIB is identified by a unique address. The instru-

ment address value is a whole decimal number between 00 and 30.

Set the address using the address switch, refer to figure 8, located on the rear panel.

The switch is binary coded and comes factory set to address 24 (1-1000). To change

the address, slide the switches with a small tool or pencil. Switch positions six and

seven are not currently used and should be left at zero. Switch eight is preset at the

factory and should not be changed.

Preliminary

Equipment Check

Check for proper equipment operation using the performance test procedures in

Chapter 5 -Maintenance. If the unit does not pass the performance test, notify Bird

Electronic Corporation.

11

Installation

Figure 7

GPIB Connector

)(

)(

)(

)(

)(

)(

)(

)(

Code

Binary

0

1

Switch

Number:

Value:

Binary

1

2

3

4

5

6

7

8

1

2

4

8

16

Not Used

Not Used

(Factory

Preset)

ON

OFF

Figure 8

Address

Select Switch