Fill the coolant reservoir according to the following instructions.

WARNING

Never attempt to connect or disconnect RF cables while RF power is on. Radiated RF is a

potential health hazard.

WARNING

Coolant containing ethylene glycol is a potential health hazard. Avoid ingestion, inhaling of

vapors and eye and skin contact.

CAUTION

The Calorimeter can overheat if operated without sufficient coolant. Failure to keep coolant

reservoir adequately filled can result in serious damage to the equipment.

CAUTION

Use only premixed coolant Bird part number 6091-120. Use of other solutions will damage

the instrument and void all warranties.

CAUTION

Do not apply RF power to the load without proper coolant flow. Failure to keep coolant at

proper flow rate can result in serious damage to the equipment.

The Calorimeter requires 525 ml (10 fluid ounces) of coolant. Fill the coolant reser-

voir as follows. Refer to figure 13.

1. Remove the power cord from the Calorimeter.

2. Remove the top cover from the Calorimeter case (if supplied).

23

Maintenance

Figure 13

Filling Coolant

Reservoir