Tabella a-3 el, A-12 – Physio-Control LIFEPAK 15 Manuale d'uso

Pagina 230

Advertising
background image

A-12

Monitor/defibrillatore LIFEPAK 15 - Istruzioni operative

Tabella A-3 Limiti degli allarmi

SEGNO VITALE
(VS)

VALORE VS
PAZIENTE

LIMITI AMPI*

LIMITI STRETTI*

GAMMA DEI LIMITI

LIMITI PREDEFINITI**

INFERIORE SUPERIORE INFERIORE SUPERIORE INFERIORE SUPERIORE INFERIORE

SUPERIORE

Frequenza
cardiaca (FC)

Frequenza
polso (FP)

<60

-20

+35

-10

+25

30–150 100–250 50

150

60–79

-25

+40

-20

+30

80–104

-30

+40

-30

+30

(BPM)

>105

-35

+45

-25

+25

SpO

2

(%)

>90

-5

+3

-5

+3

50

90–100

85

100

<90

-5

+3

-5

+3

PS sistolica
(mmHg)

<90

-20

+35

-10

+25

30

245

50

200

90–114

-20

+35

-10

+25

115–140 -25

+35

-10

+20

>140

-25

+35

-10

+20

PS diastolica
(mmHg)

<65

-15

+25

-10

+25

12

210

20

150

65-90

-15

+15

-15

+10

>90

-15

+15

-15

+10

EtCO

2

(mmHg/%)

>40/5,3

-10/-1,3 +15/+2,0 -10/-1,3 +15/+2,0 5/0,7

70/9,2

15/2,0

50/6,6

≤40/5,3

-10/-1,3 +15/+2,0 -10/-1,3 +15/+2,0

CO

2

inspirata

(mmHg/%)

n.a.

+5/+0,7

n.a.

+3/+0,4

n.a.

0/0–10/
1,3

8/1.1

Frequenza

respiratoria (FR)

<15

-8

+8

-4

+4

5-15

10–60

5

30

≥15

-15

+15

-8

+8

PA sistolica
(mmHg)

<15

-6

+12

-4

+6

10

100

10

40

15–35

-8

+16

-6

+8

>35

-12

+16

-8

+10

PA diastolica
(mmHg)

<5

-4

+12

-4

+8

0

50

0

18

5–13

-4

+16

-6

+6

>13

-6

+16

-6

+6

PVC (mmHg)

≥9

-10

+10

-5

+5

0

25

0

15

ICP, LAP
(mmHg)

<15

-6

+6

-4

+4

0

40

0

18

≥15

-6

+8

-4

+6

*Quando si impostano gli allarmi, i numeri indicati vanno aggiunti ai valori dei segni vitali del paziente o sottratti da essi.

**I limiti predefiniti vengono stabiliti quando gli allarmi sono impostati come attivi.

Advertising