Chapter 4

Operating Instructions

WARNING

Never attempt to connect or disconnect equipment

from the transmission line while RF power is being

applied. Leaking RF energy is a potential health

hazard.

Operating

Precautions

Before operating the equipment, perform the following

steps:

1. Carefully check the condition of the

RF cable and connectors used in the

circuitry.

2. The RF cable lengths, adapters and

connectors should be kept to a mini-

mum.

3. When other than 50 ohm connectors

are used, keep at the transmitter jack

only. This will keep the RF cable free

from standing waves.

Operator’s

Test for

Accuracy

Operating checks should consist mainly of comparison be-

tween this wattmeter and a Thruline Wattmeter or other

wattmeters of the same type. This will indicate an accuracy

of within ± 10 percent full scale due to the ± 5 percent full

scale limits of each instrument. Repeat the cross-check sev-

eral times to eliminate possible error from transmitter

variations.

Frequency

Response

These wattmeters are broad band instruments. Misreading

can occur with the presence of harmonics or spurious fre-

quencies. To prevent misreading due to harmonics, connect

a low pass filter between the transmitter and wattmeter. Do

not continue operation of the wattmeter when large spuri-

ous responses are present.

Zero Adjust

If the meter pointer does not read zero under no-load condi-

tions, turn the zero adjust screw (located below the meter

face) slightly to the right or left with a small screwdriver

until the pointer is exactly on the zero mark. Refer to figure

4.

9