10 technical data, 1 power supply, 2 ambient conditions – Eppendorf 5430 R Centrifuge Manuale d'uso

Pagina 59: Power supply, Ambient conditions

Advertising
background image

Technical data

Centrifuge 5430/5430 R

English (EN)

59

10

Technical data

10.1

Power supply

Centrifuge 5430

Centrifuge 5430 R

10.2

Ambient conditions

Mains connection

230 V, 50 to 60 Hz
120 V, 50 to 60 Hz
100 V, 50 to 60 Hz

Current consumption

3 A (230 V)
6 A (120 V)
7 A (100 V)

Power consumption

Maximum 475 W

EMC: interference emission (radio
interference)

EN 61326 - category B

EMC: noise immunity

EN 61326

Overvoltage category

II

Protection class

1

Fuses

4.0 AT (230 V)
8.0 AT (120 V/100 V)

Mains connection

230 V, 50 to 60 Hz
120 V, 50 to 60 Hz
100 V, 50 to 60 Hz

Current consumption

6 A (230 V)
12 A (120 V/ 100 V)

Power consumption

Maximum 1050 W

EMC: Interference emission (radio
interference)

EN 61326 – Class B (230 V/120 V)
EN 61326 – Class A (100 V)

EMC: Noise immunity

EN 61326

Overvoltage category

II

Protection class

1

Fuses

Thermal overcurrent protective switch 7 A (230 V)
Thermal overcurrent protective switch 15 A (120 V/100 V)

Ambience

For indoor use only.

Ambient temperature

Centrifuge 5430: 4 to 40°C
Centrifuge 5430 R: 10 to 40°C

Maximum relative humidity

75 %, non-condensing humidity

Atmospheric pressure

Use up to an altitude of 2000 m above MSL.

Degree of pollution

2

Advertising
Questo manuale è associato alle seguenti prodotti: