1 rcf display and calculation, Rcf display and calculation – Eppendorf 5430 R Centrifuge Manuale d'uso

Pagina 16

Advertising
background image

Product description
Centrifuge 5430/5430 R
English (EN)

16

2.4.1

rcf display and calculation

Use the

rpm/rcf key to switch the speed of centrifugation display from rpm to g-force

(rcf). Ensure that the g-force displayed during switching is standardized to suit the
rotor in question without an adapter. When adapters are used, the following maximum
g-forces (rcf) can be achieved at maximum speed:

Rotor

Adapter

Max.centrifugation
radius r

max

[cm]

Max. g-force (rcf)

Rotor FA-45-48-11/
Rotor F-45-48-11

Without adapter

Outer ring: 10.1
Inner ring: 8.9

Outer ring: 18,210
Inner ring: 16,048

For 0.2 mL PCR tubes

Outer ring: 8
Inner ring: 6.8

Outer ring: 14,425
Inner ring: 12,261

For 0.4 mL tubes

Outer ring: 10.1
Inner ring: 8.9

Outer ring: 18,210
Inner ring: 16,048

For 0.5 mL tubes

Outer ring: 9
Inner ring: 7.8

Outer ring: 16,229
Inner ring: 14,065

For 0.6 mL Microtainer

Outer ring: 10.1
Inner ring: 8.9

Outer ring: 18,210
Inner ring: 16,048

Rotor FA-45-30-11/
Rotor F-45-30-11

Without adapter

9.5

20,871

For 0.2 mL PCR tubes

7.4

16.215

For 0.4 mL tubes

9.5

20,871

For 0.5 mL tubes

8.4

18,407

For 0.6 mL Microtainer

9.5

20,817

Rotor
FA-45-24-11-Kit

Without adapter

9.8

19,090

For 0.2 mL PCR tubes

7.7

15,000

For 0.4 mL tubes

9.8

19,090

For 0.5 mL tubes

8.7

16,950

For 0.6 mL Microtainer

9.8

19,090

Rotor FA-45-16-17

For 5.0 mL micro test tubes

9.4

21,191

Rotor S-24-11-AT

Without adapter

8.9

16,049

Rotor
FA-45-24-11-HS

Without adapter

8.8

30,130

For 0.2 mL PCR tubes

6.7

22,940

For 0.4 mL tubes

8.8

30,130

For 0.5 mL tubes

7.7

26,364

For 0.6 mL Microtainer

8.8

30,130

Rotor
F-45-64-5-PCR

For PCR strips, inside

7.7

11,987

For PCR strips, outside

8.7

13,543

Rotor
F-45-18-17-Cryo

Without adapter

9.4

8,320

For Cryo tubes

9.0

7,970

Advertising
Questo manuale è associato alle seguenti prodotti: