Eppendorf 5430 R Centrifuge Manuale d'uso

Pagina 14

Advertising
background image

Product description
Centrifuge 5430/5430 R
English (EN)

14

Rotor
FA-45-24-11-HS

24 tubes 1.5/2.0 mL.
With adapters:

• 0.2 mL PCR tubes
• 0.4 mL tubes
• 0.5 mL tubes
• 0.6 mL Microtainer

30,130 × g/
17,500 rpm

3.75 g

• Aerosol-tight

(3)

rotor

lid (aluminum).

• Max. g-force/speed

(30,130 × g/17,500
rpm) only with tubes
approved by the man-
ufacturer for this
speed.

• PTFE-coated (particu-

larly resistant to
chemicals), marked:
coated

• Spin columns avail-

able, better with rotor
FA-45-24-11-kit.

• The rotor can only be

tightened and loos-
ened using the special
rotor key for rotor
FA-45-24-11-HS (see
p. 42).

≤ 21 s/16 s
≤ 61 s/65 s (soft)

Rotor
F-45-64-5-PCR

64 PCR tubes (0.2 mL)
or
eight 5-tube or 8-tube
PCR strips,
each with the enclosed
adapters.

13,543 × g/
11,800 rpm

3.4 g
(without
adapter)

≤ 12 s/15 s
≤ 62 s/65 s (soft)

Rotor
F-45-18-17-Cryo

18 Cryo tubes
or
18 sealable centrifugation
tubes,
max. Ø: 16.9 mm.
With supplied adapters: max.
Ø: 13.4 mm, max. tube length:
50 mm.

8,324 × g/
8,900 rpm

8.7 g

• Setting the g-force/

speed in increments
of 10 × g or 10 rpm.

≤ 8 s/11 s
≤ 77 s/85 s (soft)

Max. capacity

Max. g-force
(rcf)/speed
(rpm) without
adapter

Max.
load per
rotor
bore

(1)

Notes

Acceleration/
deceleration time

(2)

(soft): with soft ramp

Advertising
Questo manuale è associato alle seguenti prodotti: