4 removing the rotor, 7 standby mode, Removing the rotor – Eppendorf 5430 R Centrifuge Manuale d'uso

Pagina 41: Standby mode, See p. 41)

Advertising
background image

Operation

Centrifuge 5430/5430 R

English (EN)

41

5.6.4

Removing the rotor

1.

Turn the rotor nut

counterclockwise using the supplied rotor key.

Rotor FA-45-24-11-HS: use special rotor key.

2.

Remove the rotor vertically in an upward motion.

3.

Only 5430 R: Switch off the centrifuge after use and empty the condensation water tray (remove it
from the left side of the device). Leave centrifuge lid fully opened and protect it against closing.

5.7

Standby mode

The centrifuge automatically changes from the ready state to the standby mode if the following
prerequisites are met:

• The centrifuge was not used for the time set in the device menu (1 to 60 min)(see Settings on p. 49)

Only 5430 R: The centrifuge lid is open.

In the

Standby mode, the following applies :

• The standby key

lights red.

Only 5430 R: The rotor chamber is not cooled (see Continuous cooling on p. 38).

In the

Ready state, the following applies:

• The centrifugation parameters are displayed.

• The standby key

lights green.

Only 5430 R: The rotor chamber is cooled when the centrifuge lid is closed (see Continuous cooling on
p. 38)
.

You can switch between standby mode and ready state at any time when centrifugation is not performed by
pressing the standby key.

 Swing-bucket rotors: remove the buckets before inserting and/or removing the rotor. Use

both hands to pick up the rotor cross.

 Rotor F-35-6-30: only use the rotor removal tool supplied to insert or remove the rotor.

Advertising
Questo manuale è associato alle seguenti prodotti: