Installation
Centrifuge 5430/5430 R
English (EN)

26

4.3

Installing the instrument

Prerequisites

The device is on a suitable lab bench.

Perform the following steps in the sequence described.

1.

Let the device warm up to ambient temperature for at least 4 hours to avoid damaging electronic
assemblies as a result of condensation and avoid damaging the compressor (only 5430 R).

2.

Check that the mains voltage and frequency match the requirements on the device type plate.

3.

Connect the centrifuge to the mains and switch it on using the mains/power switch on the rear of the
device (Centrifuge 5430) or at the right side of the device (Centrifuge 5430 R).

The standby key

lights green.

The display is active.

Only 5430: Lid opens automatically

4.

Only 5430: Remove the transport securing device of the motor shaft.

5.

If the scope of delivery includes a rotor, dismantle and remove it using the supplied rotor key.

6.

Only 5430: Remove the transport securing device of the air guide ring.

NOTICE! Damage to electronic components due to condensation.
Condensate can form in the device after it has been moved from a cool environment to a
warmer environment.
 After installing the device, wait for at least 4 h. Only then connect the device to the mains

power supply.

NOTICE! Centrifuge 5430 R: compressor damage after improper transport.
 Only switch on the centrifuge 4 h after installation.