Cleaning

The most important cleaning task is to remove accumula-

tions of dust, lint, and grime that could interfere with

proper air circulation and therefore efficient heat transfer.

The cooling fins should be checked periodically for possible

collection of dust and lint. When cleaning is necessary, fol-

low these instructions.

Preparation:

s

Material: Lint-free cleaning cloth

s

Mild detergent

Outside

Surfaces

The outside surface of the instrument should be wiped free

of dust and dirt when necessary. Excessive collection of dust

and lint on the cooling fins will interfere with the efficient

dissipation of heat. Clean the RF input connector with a

self-drying, non-residue forming contact cleaner. Take spe-

cial care to clean the metallic contact surface and the ex-

posed faces of the teflon insulator.

It is particularly important to keep the following areas

clean:

s

Mating surfaces of the socket and plug-in ele-

ment including; the bore of the socket, and the

circumference of the element body.

s

Bottom rim of the element body and the seat

of the socket in the line section.

s

The ends of the insulated dc contacts on the

element. The ends should be clean and

smooth.

Inspection

Periodically perform the following checks:

s

Make sure all connections are clean and tight.

s

Disconnect the RF power and check the zero

setting without power, adjust if necessary.

s

Inspect around the clamping band of the con-

nector and also around the thermoswitch, for

signs of possible coolant leakage.

s

Check the connector assembly for tightness.