Safety, 1 intended use, 2 user profile – Eppendorf 5424 R Centrifuge Manuale d'uso

Pagina 12: 3 application limits, 2 maximum service life for accessories, Safety 3.1, Intended use, User profile, Application limits 3.3.1, Maximum service life for accessories

Advertising
background image

3 Safety

Centrifuge 5424 / 5424 R — Operating manual

12

3

Safety

Operating manual

Centrifuge 5424 / 5424 R

see

on p.

Fig.

Tab.

p.

3.1

Intended use

The Centrifuge 5424 / 5424 R is intended exclusively for indoor use and for separating aqueous
solutions and suspensions of various densities in approved test tubes.

3.2

User profile

This device may only be operated by correspondingly trained specialist staff. This staff must have
carefully read the operating manual and be familiar with the function of the device.

3.3

Application limits

3.3.1

Declaration concerning the ATEX directive (94/9/EC)

Due to its design and the environmental conditions on the inside of the device, the Centrifuge
5424 / 5424 R is not suitable for use in a potentially explosive atmosphere.
The device must therefore only be used in a safe environment, such as in the open environment
of a ventilated laboratory or an extractor hood. The use of substances which may contribute to a
potentially explosive atmosphere is not permitted. The final decision with regard to the risks
connected with the use of such substances is the responsibility of the user.

3.3.2

Maximum service life for accessories

CAUTION!

Poor safety due to incorrect accessories and spare parts.
The use of accessories and spare parts other than those recommended by Eppendorf may
impair the safety, functioning and precision of the device. Eppendorf cannot be held liable for any
damage resulting from the use of non-recommended accessories and spare parts or from the
improper use of such equipment.

Only use accessories and original spare parts recommended by Eppendorf.

DANGER!

Danger of explosion!

Do not operate the device in areas where work is being carried out with explosive substances.

Do not use this device to process any explosive, radioactive or highly reactive substances.

Do not use this device to process any substances which could create an explosive
atmosphere.

CAUTION!

Risk of injury from chemically or mechanically damaged accessories.
Even small scratches and cracks can lead to serious internal material damage.

Protect all parts from mechanical damage.

Check accessories regularly.

Do not use rotors or buckets with signs of corrosion or mechanical damage (e.g.
deformations).

Do not use accessories whose maximum service life has been exceeded.

Accessory

Maximum service life from first
commissioning

Transparent polypropylene rotor lids

3 years

Plastic adapters

1 year

EN

Oper

at

ing man

u

al

Advertising