Troubleshooting
Centrifuge 5427 R
English (EN)

42

7.2

Error messages

Proceed as follows if the following error messages appear:

1. Remove fault (see remedy).
2. If necessary, repeat centrifugation.

Symptom/message Cause

Remedy

ERR 1

Rotor not detected.

 Check rotor.
 If this error message appears again,

test with a different rotor.

Err 2

Electronics fault.

 Switch centrifuge off and back on

again after > 20 s.

NO RPM (ERR 3)

Error in speed measuring system.

 Leave the device switched on until the

error message disappears (10 s or
6 min).

Err 5

Prohibited opening of lid or lid switch is
defective during a run.

 Wait for rotor to stop.

Err 6

• Error in the drive electronics.
• Drive overheated.

 Repeat the run.
 If this error message appears again,

switch centrifuge off and back on
again after > 20 s.

 Allow the drive to cool down for at

least 15 min.

Err 7

Major deviation in the speed control.

1. Wait for rotor to stop.
2. Tighten the rotor.

Err 8

Drive fault.

1. Wait for rotor to stop.
2. Repeat the run.

ERR 9

to

ERR 14

Electronics error.

 Switch centrifuge off and back on

again after > 20 s.

IMBAL (ERR 15)

Rotor is asymmetrically loaded.

 Load the rotor symmetrically and

balance.

ERR 16

to

ERR 17

Electronics error.

 Switch centrifuge off and back on

again after > 20 s.

Err 18

Temperature deviation from set value in
the rotor chamber: ±5 °C.

 Check the settings.
 Check unhindered air circulation

through the air slots.

 Thaw ice or switch off device and

allow it to cool down.

Err 19

Condenser overheated.

 Check unhindered air circulation

through the air slots.

 Allow device to cool down.

Err 20

Temperature sensor in rotor chamber is
faulty.

 Switch centrifuge off and back on

again after > 20 s.

Err 21

Temperature sensor on condenser is
faulty.

 Switch centrifuge off and back on

again after > 20 s.