Operation
Centrifuge 5427 R
English (EN)

32

5.5.5

Continuous cooling

Continuous cooling

Continuous cooling maintains the rotor chamber at the set temperature if the rotor stops.

• During continuous cooling the display shows the set temperature.
• To prevent the rotor chamber from freezing or condensation from forming, the temperature does not go

below 4 °C , irrespective of the set temperature.

• If the rotor stops, temperature control is slower than during centrifugation or a temperature control run.

ECO shut-off

Continuous cooling is switched off if the centrifuge is not used for longer than the preset time. The
centrifuge switches to standby mode.

• Default setting: Continuous cooling ends after 8 h.
• Continuous cooling can be limited to 1 h, 2 h or 4 h.
• ECO shut-off can be switched off (continuous cooling set to endless operation).

Limit continuous cooling to 1 h (2 h, 4 h).

1. Press the menu/enter key.

2. Select

TEMP

using the arrow key. Confirm with the menu/enter key.

3. Select

1 h

(

2 h, 4 h

) using the arrow key. Confirm with the menu/enter key.

4. To activate ECO shut-off after 1 h, 2 h or 4 h, open and close again the centrifuge lid after the end of the

run.

The device switches to standby mode after 1 h, 2 h or 4 h. The display shows

EP

.

5.5.6

Endless operation of continuous cooling

The ECO shut-off function can be switched off. Continuous cooling is changed to endless operation.

• Endless operation can shorten the service life of the compressor.
• The rotor chamber may freeze.

1. Press the menu/enter key.

2. Select

TEMP

using the arrow key. Confirm with the menu/enter key.

3. Select

oo

using the arrow key. Confirm with the menu/enter key.

Ending continuous cooling

4. Open the centrifuge lid to end continuous cooling.

• If the centrifuge lid is not opened and closed again after the run, ECO shut-off engages

after 8 h (default setting).

• Open the centrifuge lid to end continuous cooling early.