3 temperature monitoring, 4 start fasttemp, Temperature monitoring – Eppendorf Centrifuge 5427 R Manuale d'uso

Pagina 31: Start fasttemp, See start fasttemp on p. 31)

Advertising
background image

Operation

Centrifuge 5427 R

English (EN)

31

5.5.3

Temperature monitoring

After the set temperature has been reached, the centrifuge reacts to temperature deviations during
centrifugation as follows:

5.5.4

Start FastTemp

With the FastTemp function, you can immediately start a temperature run without samples, at rotor-specific
or temperature-specific speeds. This will quickly bring the rotor chamber, including rotor and adapter, up
to the set target temperature.

Prerequisites

• The centrifuge is switched on.
• Rotor and rotor lid are correctly mounted.
• The centrifuge lid is closed.
• The temperature and g-force (rcf) or speed (rpm) have been set for the upcoming centrifugation (see

Centrifuging on p. 33).

1. Press the fast temp key.

The display shows

FT

Actual temperature in the rotor chamber

g-force (rcf) or speed (rpm)

The temperature control run FastTemp automatically ends when the target temperature has been
reached. A periodic signal tone sounds.

2. Press the start/stop key to end the temperature control run early.

Deviation from target temperature

Action

±3 °C

Temperature display flashes.

±5 °C

Display shows Err 18. Centrifugation is stopped automatically.

• The centrifuge only stops the run once the rotor has reached the set temperature.

Therefore, there may be a delay of approx. 15 min (rotor S-24-11-AT: 30 min) between the
display indicating the set temperature has been reached and the automatic end of the
temperature control run.

Advertising