En - 5 4. operating instructions, Instrument description 4.1.1. control description – HT instruments QUICKLAN6050 Manuale d'uso

Pagina 24

Advertising
background image

QuickLAN

6050

EN - 5

4. OPERATING INSTRUCTIONS

4.1. INSTRUMENT

DESCRIPTION

4.1.1. Control

description

1

9

7

5

2

3

8

TEST

nF

ft

m

11

7

6

4

1

10

2

3

4

12

Fig. 2: Instrument and remote identifiers #1 ~ #4

LEGEND:

1. LCD Display
2. TEST key
3. Power key
4. SET key
5. LENGTH measurement key
6. WIREMAP test key
7. ▼ ▲keys
8.

↵ ENTER key

9. M (Memory) R (Read) key
10. RJ45 jack
11. Remote Identifiers #1 ~ #4

(Identifiers #2, #3, #4 optional)

12. BNC connector


4.2. FAILURES

DETECTED

4.2.1. SHORTS
Range of short detection:

0 to 350m (0 to 999ft)

Accuracy of distance to a short:
UTP/FTP:

7% + 3m (7% + 10ft)

Coaxial Cables:

10% + 10m (10% + 30ft)

Assumes short is 0Ω.

4.2.2. OPENS
Range of open detection:

0 to 350m (0 to 999ft)

Accuracy of distance to an open : UTP/FTP:

10% + 1m (10% + 3 ft)


4.2.3. SPLIT PAIRS
Range of detection:

2 to 350m (6 to 999ft)

Split pair part of the cable must be at least 2 meters (6ft) in length and greater than 10% of
the total cable length.

4.2.4. COAX

termination

measurements

Any loop resistance value between 5 and 350Ω is interpreted as a termination resistance.
Resistance value below 5Ω is considered shorts and resistance value greater than 350Ω is
not displayed.

Advertising