13.Install the top, right-hand panel with screws and lockwashers previ-

ously removed.

CAUTION

Use only premixed coolant Bird part number 6091-120. Use of other solutions will damage

the instrument and void all warranties.

14.Refill the coolant tank with fresh coolant per the instructions.

15.Slide the calorimeter unit into the component case or equipment rack

and secure with panel mounting screws removed earlier.

Checking Voltage Selector and Power Line Fuses

Follow the steps below and refer to figure 17.

1. Unplug the ac power cord from the power entry module.

2. Use a small flat blade screwdriver to gently pry the fuse drawer out of

the power entry module.

3. Slide the voltage selector card out of the fuse drawer. Reinsert the

card so that the correct line voltage appears in the window.

4. Check the current rating of the fuses with the requirements in table 2

on page 8. Install the correctly rated fuses in the fuse drawer.

5. Reinstall the fuse drawer in the power entry module.

27

Maintenance

100V-120V

200V-240V

100V-120V

WITH

250V

USE

ONL

Y

FUSES

Module

ON/OFF

Switch

Fuse

Fuse

Voltage

Selector Card

Fuse Drawer

Figure 17