Checking voltage selector and power line fuses – Bird Technologies 6091P Manuale d'uso

Pagina 37

Advertising
background image

Air Filter Cleaning

Remove and clean the air filter element according to the following instructions. Refer

to figure 14.

1. Use the flat blade of a small screwdriver to gently pry off the air filter

cover.

2. Remove the four screws and lock washers that secure the air filter to

the front panel.

3. Scrub the air filter with hot soapy water until clean. Rinse the filter

in a stream of hot water to remove all traces of soap.

4. Reinstall the filter on the front panel using screws and lockwashers

removed in step 2.

5. Snap the air filter cover back in place on the front panel.

Drain the Coolant and Replace Coolant Filter

Remove and replace the coolant filter according to the following instructions. Refer to

figures 15 and 16.

WARNING

Coolant containing ethylene glycol is a potential health hazard. Avoid ingestion, inhaling of

vapors and eye and skin contact.

1. Shut down calorimeter unit according to the shutdown procedure.

2. Remove the four screws securing the top cover to the instrument case,

if used. Place the Calorimeter unit on a flat surface with the rear

panel extending out over the edge.

3. Place a suitable container below the drain tube. With the container in

place, remove the reservoir filler cap.

4. Open the drain valve by turning the hex counter-clockwise. Allow all

of the coolant to drain from the Calorimeter.

25

Maintenance

Figure 15

Draining the Coolant

Advertising
Questo manuale è associato alle seguenti prodotti: