5

Descrizione dettagliata degli elementi di comando e funzionamento

5

Descrizione dettagliata degli elementi di comando e funzionamento

5.1

Controls and function

Operating controls are located on a control panel mounted in the door of the upright freezers, and
on the console on the right side of chest freezers.
Every New Brunswick freezer is equipped with S.M.A.R.T. Plus™ diagnostic software, to help
identify the cause of a fault or setpoint variance.
This section describes the controls and function of the control panel, (vedi Fig. 3 a pag. 16).

Abb. 3:

Display panel & keypad

Fig. 3:

Display panel & keypad

Temperature °C Display

High alarm/low alarm

TEMPERATURE

1

7

8

9

0

E

C

2

3

4

5

6

REMOTE-CONTROL

TEMP-ALARM

FILTER-CLEAN BATTERY-LOW

POWER-FAIL

FAULT

SET

TEMP

LOW

ALARM

LOCK

HIGH

ALARM

CODE

CHANGE

ALARM

TEST/

MUTE

C

1

12

8

13

15

14

11

10

9

2

7

5

4

3

6

Item

Name

Function

1

TEMPERATURE

°C

The digital display normally shows the current internal
temperature of the freezer. Temperature is displayed in 1°C
increments.

Item

Name

Function

2 & 8

[HIGH/LOW]

TEMP-ALARM

The TEMP-ALARM illuminates and the audible alarm sounds if
the freezer’s internal temperature is above/below the
user-selected alarm setpoints. The LED illuminates as soon as
the setpoint is passed. The high temperature audible alarm has
a programmable delay (vedi Impostazione del ritardo di allarme
a pag. 24)
. Af
ter the temperature returns to the normal range,
the TEMP-ALARM switches off and the audible warning stops.

Nota!

The audible alarm can be silenced by pressing the ALARM TEST/MUTE KEY. If the temperature
has not returned to normal range after the programmed time period, the audible warning will
sound again. This pattern will continue to repeat until the temperature returns to normal.