Maintenance

Centrifuge 5418 / 5418 R

English (EN)

29

5. Rinse the rotors and accessories thoroughly with distilled water. Rinse the rotor bores of fixed-angle

rotors particularly thoroughly.

6. Place rotors and accessories on a cloth to dry. Place fixed-angle rotors with the rotor bores facing

downwards to allow the bores to also dry.

7. Correctly reinsert the rotor lid sealing ring in the clean and dry groove.

8. Clean the rotor cone with a soft, dry and lint-free cloth. Do not lubricate the rotor cone.

9. Inspect the rotor cone for damage.

10. Place the dry rotor onto the motor shaft.

11. Tighten the rotor nut firmly by turning it

clockwise with the rotor key.

12. Leave the rotor lid open when the rotor is not being used.

6.3

Additional service instructions for Centrifuge 5418 R

Empty and clean the condensation water tray regularly and especially after liquid spillage in the rotor
chamber. Pull out the condensation water tray from the left or right side of the centrifuge.

Regularly free the rotor chamber ice formations via thawing, by leaving the centrifuge lid open or
carrying out a short temperature control run at approx. 30

°C.

Leave the centrifuge lid open when not in use for a long period.

Wipe up condensate in the rotor chamber using a soft, absorbent cloth.

Remove dust deposits from the ventilation slits of the centrifuge using a brush or swab at the latest
every six months. First switch off the device and remove the power plug.

6.4

Glass breakage

When using glass tubes there is a risk of glass breakage in the rotor chamber. The resulting glass splinters
are swirled around in the rotor chamber during centrifugation and have a sandblasting effect on the rotor
and accessories. The smallest glass particles become lodged in the rubber parts (e.g., the motor guide, the
rotor chamber gasket, and the rubber mats of adapters).

Do not immerse the rotor in liquid as liquid can get trapped inside the cavities.