Set number of floats, Set baud rate, Set receive to transmit delay – VEGA VEGAFLEX 81 Modbus and Levelmaster protocol Manuale d'uso

Pagina 91

Advertising
background image

91

11 Appendice

VEGAFLEX 81 • Protocollo Modbus e LevelMaster

41828-IT

-130416

Response:

Parametri

Length

Code/Data

Assign Unit Number

6 characters ASCII

UuuNOK
uu = new Address

Set number of Floats

Request:

Parametri

Length

Code/Data

Set number of Floats

5 characters ASCII

UuuFn

Response:

Parametri

Length

Code/Data

Set number of Floats

6 characters ASCII

UuuFOK

Se il numero viene impostato su 0, non viene più segnalato alcun livello

Set Baud Rate

Request:

Parametri

Length

Code/Data

Set Baud Rate

8 (12) characters ASCII

UuuBbbbb[b][pds]
Bbbbb[b] = 1200, 9600 (default)
pds = parity, data length, stop bit

(optional)
parity: none = 81, even = 71 (defau-

lt), odd = 71

Response:

Parametri

Length

Code/Data

Set Baud Rate

11 characters ASCII

Esempio: U01B9600E71
Modificare l'apparecchio all'indirizzo 1 impostando il baud rate 9600, parità pari, 7 bit di dati, 1 bit di

stop

Set Receive to Transmit Delay

Request:

Parametri

Length

Code/Data

Set Receive to Transmit Delay

7 characters ASCII

UuuRmmm
mmm = milliseconds (50 up to 250),

default = 127 ms

Response:

Parametri

Length

Code/Data

Set Receive to Transmit Delay

6 characters ASCII

UuuROK

Advertising